The Nuclear Magnetic Flowmeter for Monitoring the Consumption and Composition of Oil and Its Complex Mixtures in Real-Time
March 1, 2023
Abstract
The necessity of increasing the efficiency of primary oil purification in a drilling station or an offshore platform has been substantiated. We consider the problems that arise during the primary processing of oil mixtures. Important conditions for increasing the efficiency of primary purification (separation) of oil mixtures include measuring the consumption and determining the content of various impurities (water, undissolved particles) and air in them, with an error of no more than 2%. We analyzed the possibilities of using various designs of flowmeters to measure the consumption of the oil mixture coming from a well. It is also necessary to use other measuring instruments to control the state of this mixture, which creates additional problems (searching for an appropriate locations to place them, providing the required operating conditions). Various designs of nuclear magnetic flowmeters−relaxometers were considered, making it possible to measure the consumption of a liquid medium and its times of longitudinal T1 and transverse T2 relaxation with one device. The measured values of T1 and T2 determine the state of the medium. The design of the industrial nuclear magnetic flowmeter−relaxometer M-Phase 5000, which is used to control the flow and quality of oil and oil products, was considered in more detail. Problems were identified that did not allow using this design of a nuclear magnetic flowmeter−relaxometer in a drilling rig or offshore platform. A new design of a nuclear magnetic flowmeter−relaxometer was developed, implementing the methods for measuring q, T1, and T2. These methods and various technical solutions make it possible to use this device at a drilling station or offshore platform. The measurement errors of the consumption q, T1, and T2 were determined. The results of various media studies are presented and compared with q, T1, and T2 measurements on other devices and measured volume (to confirm the adequacy of q measurements). The application scopes of the developed nuclear magnetic flowmeter−relaxometer were determined, in addition to the systems of primary oil processing.
The necessity of increasing the efficiency of primary oil purification in a drilling station or an offshore platform has been substantiated. We consider the problems that arise during the primary processing of oil mixtures. Important conditions for increasing the efficiency of primary purification (separation) of oil mixtures include measuring the consumption and determining the content of various impurities (water, undissolved particles) and air in them, with an error of no more than 2%. We analyzed the possibilities of using various designs of flowmeters to measure the consumption of the oil mixture coming from a well. It is also necessary to use other measuring instruments to control the state of this mixture, which creates additional problems (searching for an appropriate locations to place them, providing the required operating conditions). Various designs of nuclear magnetic flowmeters−relaxometers were considered, making it possible to measure the consumption of a liquid medium and its times of longitudinal T1 and transverse T2 relaxation with one device. The measured values of T1 and T2 determine the state of the medium. The design of the industrial nuclear magnetic flowmeter−relaxometer M-Phase 5000, which is used to control the flow and quality of oil and oil products, was considered in more detail. Problems were identified that did not allow using this design of a nuclear magnetic flowmeter−relaxometer in a drilling rig or offshore platform. A new design of a nuclear magnetic flowmeter−relaxometer was developed, implementing the methods for measuring q, T1, and T2. These methods and various technical solutions make it possible to use this device at a drilling station or offshore platform. The measurement errors of the consumption q, T1, and T2 were determined. The results of various media studies are presented and compared with q, T1, and T2 measurements on other devices and measured volume (to confirm the adequacy of q measurements). The application scopes of the developed nuclear magnetic flowmeter−relaxometer were determined, in addition to the systems of primary oil processing.
Record ID
Keywords
consumption, flow, impurities, longitudinal T1 and transverse T2 relaxation times, measurement error, nuclear magnetic resonance, oil mixture, pipeline, purification, signal-to-noise ratio, state of the medium
Subject
Suggested Citation
Davydov R, Davydov V, Dudkin V. The Nuclear Magnetic Flowmeter for Monitoring the Consumption and Composition of Oil and Its Complex Mixtures in Real-Time. (2023). LAPSE:2023.13613v1
Author Affiliations
Davydov R: Institute of Physics and Mechanics, Peter the Great St. Petersburg Polytechnic University, 195251 St. Petersburg, Russia [ORCID]
Davydov V: Institute of Electronics and Telecommunications, Peter the Great St. Petersburg Polytechnic University, 195251 St. Petersburg, Russia; All Russian Research Institute of Phytopathology, 143050 Moscow, Russia
Dudkin V: Department of Photonics and Communication Lines, The Bonch-Bruevich Saint Petersburg State University of Telecommunication, 193232 St. Petersburg, Russia [ORCID]
Davydov V: Institute of Electronics and Telecommunications, Peter the Great St. Petersburg Polytechnic University, 195251 St. Petersburg, Russia; All Russian Research Institute of Phytopathology, 143050 Moscow, Russia
Dudkin V: Department of Photonics and Communication Lines, The Bonch-Bruevich Saint Petersburg State University of Telecommunication, 193232 St. Petersburg, Russia [ORCID]
Journal Name
Energies
Volume
15
Issue
9
First Page
3259
Year
2022
Publication Date
2022-04-29
ISSN
1996-1073
Version Comments
Original Submission
Other Meta
PII: en15093259, Publication Type: Journal Article
Record Map
Published Article

LAPSE:2023.13613v1
This Record
External Link

https://doi.org/10.3390/en15093259
Publisher Version
Download
Meta
Record Statistics
Record Views
208
Version History
[v1] (Original Submission)
Mar 1, 2023
Verified by curator on
Mar 1, 2023
This Version Number
v1
Citations
Most Recent
This Version
URL Here
https://psecommunity.org/LAPSE:2023.13613v1
Record Owner
Auto Uploader for LAPSE
Links to Related Works
(0.27 seconds)