Embedded Control in Wearable Medical Devices: Application to the Artificial Pancreas
Stamatina Zavitsanou, Ankush Chakrabarty, Eyal Dassau, Francis J. Doyle
July 30, 2018
Abstract
Significant increases in processing power, coupled with the miniaturization of processing units operating at low power levels, has motivated the embedding of modern control systems into medical devices. The design of such embedded decision-making strategies for medical applications is driven by multiple crucial factors, such as: (i) guaranteed safety in the presence of exogenous disturbances and unexpected system failures; (ii) constraints on computing resources; (iii) portability and longevity in terms of size and power consumption; and (iv) constraints on manufacturing and maintenance costs. Embedded control systems are especially compelling in the context of modern artificial pancreas systems (AP) used in glucose regulation for patients with type 1 diabetes mellitus (T1DM). Herein, a review of potential embedded control strategies that can be leveraged in a fully-automated and portable AP is presented. Amongst competing controllers, emphasis is provided on model predictive control (MPC), since it has been established as a very promising control strategy for glucose regulation using the AP. Challenges involved in the design, implementation and validation of safety-critical embedded model predictive controllers for the AP application are discussed in detail. Additionally, the computational expenditure inherent to MPC strategies is investigated, and a comparative study of runtime performances and storage requirements among modern quadratic programming solvers is reported for a desktop environment and a prototype hardware platform.
Keywords
artificial pancreas, embedded control systems, model predictive control (MPC), safety-critical applications, software architecture
Suggested Citation
Zavitsanou S, Chakrabarty A, Dassau E, Doyle FJ. Embedded Control in Wearable Medical Devices: Application to the Artificial Pancreas. (2018). LAPSE:2018.0186
Author Affiliations
Zavitsanou S: Harvard John A. Paulson School of Engineering and Applied Sciences, Harvard University, Cambridge, MA 02138 , USA
Chakrabarty A: Harvard John A. Paulson School of Engineering and Applied Sciences, Harvard University, Cambridge, MA 02138 , USA [ORCID]
Dassau E: Harvard John A. Paulson School of Engineering and Applied Sciences, Harvard University, Cambridge, MA 02138 , USA
Doyle FJ: Harvard John A. Paulson School of Engineering and Applied Sciences, Harvard University, Cambridge, MA 02138 , USA
[Login] to see author email addresses.
Journal Name
Processes
Volume
4
Issue
4
Article Number
E35
Year
2016
Publication Date
2016-09-23
ISSN
2227-9717
Version Comments
Original Submission
Other Meta
PII: pr4040035, Publication Type: Review
Record Map
Published Article

LAPSE:2018.0186
This Record
External Link

https://doi.org/10.3390/pr4040035
Publisher Version
Download
Files
Jul 30, 2018
Main Article
License
CC BY 4.0
Meta
Record Statistics
Record Views
1335
Version History
[v1] (Original Submission)
Jul 30, 2018
 
Verified by curator on
Jul 30, 2018
This Version Number
v1
Citations
Most Recent
This Version
URL Here
http://psecommunity.org/LAPSE:2018.0186
 
Record Owner
Auto Uploader for LAPSE
Links to Related Works
Directly Related to This Work
Publisher Version
(0.08 seconds)

[0.08 s]